Description
Product spec
CE certified | Yes |
Accuracy Standards(ISO 8655) | Yes |
Autoclave | Yes |
Ordering Details
Maximum Permissible Systematic Error | Maximum Permissible Random Error | |||||||
Code | Capacity | Increment | Test Volume | Inaccuracy | Inaccuracy | Imprecision | Imprecision | No. / Case |
E11139 | 0.5 – 10 µl | 0.1 µl | 10 µl 5 µl 1 µl |
± 1.50 % ± 2.50 % ± 4.00 % |
± 0.15 µl ± 0.125 µl ± 0.04 µl |
± 1.50 % ± 2.50 % ± 4.00 % |
± 0.15 µl ± 0.125 µl ± 0.04 µl |
1 |
E11140 | 5 – 50 µl | 0.5 µl | 50 µl 25 µl 5 µl |
± 1.00 % ± 1.50 % ± 3.00 % |
± 0.50 µl ± 0.375 µl ± 0.15 µl |
± 0.50 % ± 1.00 % ± 2.00 % |
± 0.25 µl ± 0.25 µl ± 0.10 µl |
1 |
E11141 | 50 – 300 µl | 5 µl | 300 µl 150 µl 50 µl |
± 0.70 % ± 1.00 % ± 1.50 % |
± 2.10 µl ± 1.50 µl ± 0.75 µl |
± 0.25 % ± 0.50 % ± 0.80 % |
± 0.75 µl ± 0.75 µl ± 0.40 µl |
1 |
Reviews